POLARIS





 

 





VERTICAL LAMINAR AIRFLOW CABINET




Introduction
Designed for the situation where is requested the product protection from dangerous effects due to uncontrolled diffusion of air-transported contaminants from the environmental to the product during its handling. Since Polaris is not a biological safety cabinet, its use is not for handling pathogen products.



Applications
Normally used in laboratory of microbiology, virology, haematology and cell culture.
Furthermore, used in innovative fields of scientific research and above all in the manipulation of:
· Not pathogens cell culture
· Quality Control in pharmaceutical and food industries
· €Micro mechanical assembly



Definition
POLARIS 48 and 72 cabinets produced by Steril S.p.A. are defined: Vertical laminar flow cabinet as designed to fulfil the above-described application. Class 100 (M 3.5) as FS 209e with partial frontal working aperture for aspiration 15% of the total air involved, and the remaining 85% is recirculated through the HEPA filter in a working chamber with a laminar air flow velocity of 0.40 m/s +/-20%.

Main Technical Specification
Lower part of back panel and spillage tray in stainless steel AISI 304L "2B" finishing, thickness 12/10, smooth and sealed edges. Back panel in steel epoxy polyester powder painted, resistant to the most common industrial disinfectant. Working surface in stainless steel AISI 304L "2B" finishing, 4 and 6 sectors removable and autoclavable. Available both in perforated and solid version. Side walls in toughened safety glass. Front window, in toughened safety glass electrical driven. Construction structure in steel epoxy-polyester powder painted, resistant to the most common industrial disinfectants. Ventilation system by means of n°1 motor blower for POLARIS 48 dedicated to the one-way airflow in the working chamber, equal to 85% of the total involved air and to exhaust the remaining 15%. The motor blower is centrifugal type, direct driven motor with double aspiration and protection factor IP22. POLARIS 72 is supplied with n° 2 motor blowers in parallel way system Air filtration, both recirculate and exhaust air by means of HEPA "H14 type" filter with efficiency of 99,999% at 0,3 micron particles size with DOP/DOS test (as defined to EN1822).


Command and Controls
Air ventilation control by means of manual electric regulation of the revolution velocity of the motor-blower achieves via interfaced to microprocessor. It guarantees the activity, control and monitoring of air ventilation. Alphanumeric display (optional) showing Laminar air flow speed in m/s.

Standard features
· Fluorescent lights (n°1 - 36W for POLARIS 48 and 58W for POLARIS 72)
· Power socket 4 A UNEL IP 44 (internal back panel-right side)
· Vacuum line with manual tap (internal back panel - left side) manufactured with piping made of copper and hose-bard 10 mm. diameter (external) for the connection to the main line (max pressure 4 bar).
·DOP/DOS test inlet hose-barb port

Available size 120 and 180 cm.


 Trademarks are propriety of respective owners. Copyright © 2003 E.U.S.S. Ltd - WEBMASTER